HOME   |    PDF   |   


Title

Comparative cyclic fatigue resistance of heat-Treated rotary nickel-Titanium systems in curved canals: An in vitro study

 

Authors

Omkar Eswara Babu Danda 1,*, Jitendra Sharan 2, Pallawi Sinha 3, B. Nagesh 4, Thati Jyotsnanjali 5 & Tarun Kumar Singh 6

 

Affiliation

1Department of Conservative Dentistry and Endodontics, Dr NTR Institute of Health Sciences, Vijayawada, Kanuru andhra Pradesh, India; 2Department of Orthodontics & Dentofacial orthopedics, AIIMS, Patrapada, Bhubaneswar, Odisha, India; 3Department of Prosthodontics, Hi-Tech Dental College and Hospital, Bhubaneswar, Odisha, India; 4Department of Conservative Dentistry and Endodontics, Sibar Institute of Dental Sciences, Takkellapadu, Guntur andhra Pradesh, India; 5Department of Conservative Dentistry and Endodontics, Anil Neerukonda Institute of Dental Sciences, Bheemunipatnam andhra Pradesh, India; 6Department of Conservative Dentistry and Endodontics, AIIMS, Bathinda, Punjab, India; *Corresponding author

 

Email

Omkar Eswara Babu Danda - E-mail: g3eswar@gmail.com; Phone:+91 8074762630
Jitendra Sharan - E-mail: jsbmds@gmail.com; Phone: +91 9968856761
Pallawi Sinha - E-mail: palls1712@gmail.com; Phone:+91 7291958754
B. Nagesh - E-mail: drnageshendo@sids.ac.in; Phone: +91 9848885658
Thati Jyotsnanjali - E-mail: jyotsnanjali94@gmail.com; Phone: +91 8892420520
Tarun Kumar Singh - E-mail: drtarunkumarsingh@gmail.com; Phone: +91 9560627466
 

Article Type

Research Article

 

Date

Received October 1, 2026; Revised October 31, 2026; Accepted October 31, 2026, Published October 31, 2026

 

Abstract

Cyclic fatigue-related instrument separation remains a major concern during rotary nickel-titanium instrumentation of curved canals, prompting advances in alloy processing and heat treatment to improve fatigue resistance. Therefore, it is of interest to compare the cyclic fatigue resistance of Flexicon X3, Hyflex CM and Mtwo instruments in simulated canals with 60° and 90° curvatures. Hence, sixty instruments (ISO size 25, 0.06 taper, 25 mm length) were divided into three groups of 20 and tested using a dynamic fatigue model in stainless-steel canals at manufacturer-recommended speed and torque; time to fracture was recorded, the number of cycles to failure was calculated and data were analysed using two-way ANOVA followed by Tukey post-hoc tests at a significance level of p < 0.05. Flexicon X3 presented the highest cyclic fatigue resistance in both curvatures, followed by Hyflex CM and Mtwo. Significant differences were found between the groups (p < 0.05), but no significant effect of canal curvature was found. Thus, the Flexicon X3 showed higher cyclic fatigue resistance, indicating that the annealed heat treatment and metallurgical properties may improve the performance of this instrument in curved canals.

 

Keywords

Cyclic fatigue, resistance, instrument separation, endodontic instrumentation, nickel-titanium rotary instruments

 

Citation

Danda et al. Bioinformation 22(10): 7450-7454 (2026)

 

Edited by

Ashwini Dhopte

 

ISSN

0973-2063

 

Publisher

Biomedical Informatics

 

License

This is an Open Access article which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. This is distributed under the terms of the Creative Commons Attribution License.